| Approvals |
|
|
| |
NHTSA |
Listed on the Conforming Products List for Evidential Breath Measurement Devices. |
| Analytical System |
|
|
| |
Electrochemical Sensor (Fuel Cell) |
Select sensor, manufactured to Intoximeters' exacting specifications for accuracy, repeatability and long life. |
| Measurement Range |
|
|
| |
0.000 to 0.440 g/210L |
The off the shelf instrument is designed to offer a range of results that our customers commonly encounter when performing breath alcohol testing. Custom ranges, and results reported in most units of measure are available upon request. |
| Accuracy and Precision |
|
|
| |
Highly accurate and precise evidential grade results |
Intoximeters fuel cell based instruments set the industry standard for accuracy and precision. These precise and linear devices offer consistent output that results in evidentiary grade analysis. |
| Sensor Selectivity |
|
|
| |
Highly specific for alcohol |
Fuel Cell sensors are sought after for their specificity to alcohol. |
| Sampling Method |
|
|
| |
Flow, Volume, Breath Drop |
Instruments running the standard software require a consistent flow, a minimum volume and then a drop in flow before sampling a deep lung breath. Alternate sample capture algorithms are available in customized programs. |
| Testing Environment |
|
|
| |
Instrument Temperature |
The temperature of the instrument must be at, or between, 0 and 50 degrees C. |
| |
Ambient Temperature |
As long as the instrument is at temperature, the instrument can be used at, or between, ambient temperatures of -20 to 70 degrees C. |
| |
Humidity |
10% to 100% relative humidity (The instrument is not designed to be used under water.) |
| |
Ambient Pressure |
600 mbar to 1400 mbar; 450 mmHg [0 °C] to 1050 in mmHg [0 °C] |
| Mouthpiece |
|
|
| |
Style |
Split Chamber with Air Flow Diversion |
|
Installation |
Lever and set in the lighted mouthpiece guide |
| Data Input |
|
|
|
(RBT V) Keyboard |
The RBT VXL comes standard with a keyboard for ease of data entry |
|
(AS V) On board scroll buttons |
The ASVXL instrument supports alpha-numeric data entry and drop down list item selection with two scroll buttons and a data entry button on the instrument |
|
Data Entry Optional |
Data entry capability allows entry of subject name, subject ID, operator name and reason for the test |
| Visual Output |
|
|
| |
Display |
Multi-color backlit graphic LCD |
|
LED |
Side viewable LED indicates blow condition and warnings |
| Audio Output |
|
|
| |
Beeper |
Audible tone to support displayed messaging system |
| Data Output |
|
|
|
Printer |
Serial printer (thermal or impact options) |
|
Serial |
With optional serial docking station adapter, the ASVXL can download stored test data serially and, with proper software, can communicate via serial modem or through a serial Ethernet adapter |
| Data Storage |
|
|
| |
Number of Tests |
More than 7,000 test memory for a basic test record |
| Power Options |
|
|
| |
(AS V) Disposable Battery |
4 x 1.5V AA alkaline batteries |
|
(AS V) Rechargeable Battery |
4 x AA NiMH rechargeable cells |
|
Printer Battery |
Depending upon selected printer, a rechargeable NiMH or Li-Ion battery (comes with charger) |
|
Printer 12V Power Source |
Optional 12V cigarette adapter to charge printer |
| Case Construction |
|
|
| |
AS V case ABS |
Rugged casing with rubber over-molding protects instrument and internal components |
|
Nylon Carrying Case |
The RBT VXL instrument comes in a light weight blue nylon carrying case which houses the ASVXL, docking station, printer, and storage of additional dry gas calibration equipment |
| Size, Shape and Weight |
|
|
| |
Dimensions |
|
| |
Weight |
|
| Measurement Types |
|
|
| |
Automatic Direct Breath |
Auto sampling of subjects' deep lung breath samples |
| |
Manual Direct Breath |
Manual sampling option for subjects who cannot meet the minimum flow and volume requirements of the instrument. |
|
Passive Testing |
Manual or automatic sampling capabilities for passive alcohol detection |
|
Headspace Sampling |
Capable of measuring the headspace over a liquid to determine if alcohol might be present in the solution |
| Security |
|
|
|
Password |
|
| Options |
|
|
| |
Barometer |
Optional onboard barometer simplifies the use of dry gas standards |
|
Software |
Both AS V Accelerator download software and IntoxNet interface |
| |
Ethernet |
Serial to Ethernet adapters offer internet connectivity. Optionally, the ASVXL offers software that transmits encrypted data and thus provides security when sending test records over the public internet. |